Can vacuum assisted venous drainage be achieved using a roller pump in an emergency? A pilot study using neonatal
1University of Nebraska Medical Center, Omaha, Nebraska 68198-5155, USA. slhill@unmc.edu
Abstract:
There has been much advancement in perfusion technology over its 50 years of progression. One of these techniques is vacuum-assisted venous drainage (VAVD). Many perfusionists augment venous drainage using VAVD, typically from a wall vacuum source. This study explores alternates to providing VAVD if the wall vacuum fails. In two porcine laboratories, approximately 36 in. of 3/16-in. tubing was connected to a sucker return port and placed into the roller head next to the arterial pump. The vacuum was monitored with a DLP pressure monitoring system (Medtronic). This system was connected to small-bore tubing and attached to a stopcock on top of the reservoir. The vacuum was regulated using another stopcock connected to a non-filtered luer lock port on top of the reservoir or by a segment of 3 x 0.25-in.-diameter tubing attached to the vent port with a c-clamp. Vacuum drainage was achieved, ranging from -18 mmHg to -71 mmHg by manipulating the stopcock or c-clamp. Changes in venous drainage were seen by volume fluctuations in the venous reservoir. The vacuum was adjusted to account for dramatic changes. Augmented venous drainage using a roller pump can be achieved successfully during cardiopulmonary bypass (CPB). This method of active drainage can be used in lieu of wall suction or during times of emergency if wall suction fails.