Related Experiment Video
Updated: Jul 7, 2026

Assessment of Plasma Coagulation on Liver Tissue in a Large Animal Model In Vivo
Published on: August 4, 2018
Optimizing bipolar electrocoagulation for endoscopic hemostasis: assessment of factors influencing energy delivery
Loren Laine1, Gary L Long, Gregory J Bakos
1Department of Medicine, University of Southern California Keck School of Medicine, Los Angeles, California 90033, USA.
Background:
Few data inform decisions on the optimal bipolar electrocoagulation (BPEC) technique.
Objectives:
To assess how technical factors influence energy delivery and coagulation.
Design:
Prospective, randomized study in experimental models: meat, live pig mesenteric arteries.
Interventions:
Standard and prototype BPEC probes were applied at varying durations (2, 10, and 20 seconds), application forces (5, 75, and 150 g), and watt settings (10, 15, and 20 W). BPEC devices were applied to arteries with 40 g versus no additional force.
Main Outcome Measurements:
For the meat model: energy delivered, impedance, coagulation and cavitation depth, and coagulation surface area. For the mesenteric arteries: hemostasis.
Results:
The energy delivered increased with duration and force (P < .001) but not with the watt setting. Impedance rose rapidly at higher watt settings (>300 ohms within approximately 5 seconds at 20 W and approximately 10 seconds at 15 W), with a coincident drop in power. Coagulation depth and surface area correlated with energy delivered (r = 0.70-0.97). Only duration was associated with the coagulation depth (P < .001); cavitation (which occurred with a standard BPEC probe) plus coagulation depth was also associated with application force (P < .001). Hemostasis of the mesenteric arteries was achieved only with 40 g of force.
Limitations:
The accuracy of these models in predicting clinical results is uncertain.
Conclusions:
Increasing BPEC duration increased the energy delivered and the coagulation, whereas increasing the watt setting did not because of a rapid rise in impedance. Optimal BPEC technique included a lower watt setting (eg, 15 W), a longer duration (eg, approximately 10-12 seconds), and tamponade of the bleeding site.