Detecting change in left ventricular ejection time during head-up tilt-induced progressive central hypovolemia using
Gregory S H Chan1, Paul M Middleton, Branko G Celler
1Biomedical Systems Laboratory, School of Electrical Engineering and Telecommunications, Graduate School of Biomedical Engineering, University of New South Wales, Sydney, Australia.
Background:
Change in cardiac preload caused by mild hypovolemia can alter left ventricular ejection time (LVET) without noticeable change in blood pressure (BP). Previously our group has explored a novel method of LVET monitoring using a noninvasive finger photoplethysmographic pulse oximetry wave form. The current study investigated the ability of photoplethysmographic pulse oximetry wave form-derived LVET (LVETp) to identify progressive central hypovolemia induced by head-up tilt and evaluated the potential use of LVETp as an early noninvasive indicator of blood loss.
Methods:
Thirteen healthy subjects underwent graded head-up tilt from 0 degrees to 80 degrees. The response of LVETp to tilt was compared with that of interbeat heart interval (RR) and BP. Least-squares linear regression analysis was performed on an intrasubject basis between various physiologic variables and sine of the tilt angle (which is associated with the decrease in central blood volume).
Results:
During graded tilt, LVETp had a very strong negative linear correlation with sine of the tilt angle, with correlation coefficients (r) ranging from -0.961 to -0.985. At a very mild hypovolemic state (10 degrees), there was a significant decrease in LVETp compared with baseline (0 degrees) but without a significant change in RR and BP. Gradient analysis showed that LVETp was sensitive to central volume loss at all volume states (0 degrees-80 degrees), whereas RR was only responsive at mild-to-moderate and moderate hypovolemic states (20 degrees-80 degrees) but not mild hypovolemic state (0 degrees-20 degrees).
Conclusions:
LVETp has a strong association with the change in central blood volume and may be a sensitive early marker of nonhypotensive progressive central hypovolemia. Joint interpretation of LVETp and RR trends may help to characterize the extent of blood volume loss.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...