Detection of Histoplasma capsulatum antigenuria by ultrafiltration of samples with false-negative results
Lindsey Egan1, Patricia A Connolly, Deanna Fuller
1MiraVista Diagnostics, 4444 Decatur Blvd., Suite 300, Indianapolis, IN 46241, USA. legan@miravistalabs.com
Clinical and Vaccine Immunology : CVI
|February 29, 2008
Abstract:
Patients with histoplasmosis may test falsely negative for Histoplasma capsulatum antigenuria. In some cases antigen is present at levels below the assay's detection limit, and ultrafiltration could improve sensitivity. Antigen was detected following ultrafiltration in 73.8% of falsely negative specimens versus 2% of controls. Ultrafiltration improved sensitivity with a small reduction in specificity.