[Effects of different sterilization methods on the corrosion of dental fissure bur]
Rui-Min Li1, Cheng-Lin Wang, Gui-Xiang Xu
1State Key Laboratory of Oral Diseases, Sichuan University, Chengdu 610041, China.
Objective:
The purpose was to analyze the effects of three sterilization methods (dry heat sterilization, steam sterilization, chemical sterilization) on the corrosion of dental fissure burs.
Methods:
100 dental fissure burs were distributed to 10 groups. One was control, the burs in the other 9 groups were treated by dry heat sterilization, steam sterilization, chemical sterilization with 5, 10, 15 cycles respectively. Weight method, scanning electron microscope, micro-hardness measurement were used to analyze the corrosion of dental fissure burs.
Results:
The fissure burs gained their weight with cycles of sterilization. 5, 10, 15 cycles of dry heat sterilization, 10, 15 cycles of steam sterilization and 15 cycles of chemical sterilization, the weight of fissure burs were significantly increased (P < 0.05). Scanning electron microscope showed differences on the surfaces of dental fissure burs among the three sterilization groups. After sterilization, spot or partial erosion were seen on the surface of the burs. The steam sterilization groups showed the most evident changes, followed by chemical sterilization groups and dry heat sterilization groups. X-ray energy spectrometer showed the steam sterilization groups had the largest percentage of W, followed by dry heat sterilization groups, chemical sterilization groups and control group. Fe had the opposite trends. Micro-hardness reduced after sterilization. The reduction was most clear in steam sterilization group, followed by chemical sterilization and dry heat sterilization ( P< 0.05). The difference between 5 and 10 times of steam sterilization and 5, 10, 15 times of chemical sterilization were significant difference (P < 0.05). There was no significant difference between 5, 10, 15 times of dry heat sterilization (P > 0.05).
Conclusion:
The corrosion is most severe in steamsterilization group, followed by chemical sterilization, dry heat sterilization. Dry heat sterilization shows less corrosion.
More Related Videos
05:30Roughness Impact of Piezoelectric Dental Scaler on Two Distinct Flowable Composite Filling Materials
Published on: January 10, 2025
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Microbial Corrosion
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...