Comprehensive & Cost Effective Laboratory Monitoring of HIV/AIDS: an African Role Model
Published on: October 31, 2010
Affordable pediatric CD4 counting by flow cytometry in Malawi
Calman A Maclennan1, Felix Dzumani, Alinane Namarika
1Malawi-Liverpool-Wellcome Trust Clinical Research Programme, College of Medicine, University of Malawi, Blantyre, Malawi. c.maclennan@bham.ac.uk
Background:
Rapid expansion of antiretroviral therapy in Malawi has occurred in the relative absence of suitable pediatric CD4 counting facilities. We have recently validated in adults a simplified affordable flow cytometric CD4 counting method, the Blantyre count. There is a need for this technology to transfer to government laboratories run by local staff, and to be validated in children, where %CD4/lymphocyte values are required.
Methods:
We assessed agreement of %CD4/lymphocyte values determined by the Blantyre count and Panleucogate methods on an EPICS XL-MCL flow cytometer on 113 venous blood samples from HIV-seropositive children in Blantyre, Malawi. All assays were performed by two Malawian laboratory technicians.
Results:
Overall bias between the two methods was -0.13% (95% CI -0.37 to 0.11) and limits of agreement were -2.69 to 2.43% (95% CI -3.11 to -2.27 and 2.01 to 2.85). Limits of agreement were within -3.00 and 3.00 for each laboratory technician. Coefficient of variation for the Blantyre count assay was 2.0% and samples showed good stability over 5 days.
Conclusions:
The Blantyre count method can accurately determine %CD4/lymphocyte values in blood of HIV-seropositive children on an EPIC XL-MCL flow cytometer at a reagent cost of US $0.21 per test or less. The assay can be competently carried out by local laboratory technicians.