Marginal fit and microgaps of implant-abutment interface with internal anti-rotation configuration
Takuma Tsuge1, Yoshiyuki Hagiwara, Hideo Matsumura
1Nihon University Graduate School of Dentistry, 1-8-13 Kandasurugadai, Chiyoda-ku, Tokyo 101-8310, Japan.
Abstract:
To date, there is no evidence that internal anti-rotation configurations are better than external ones. As part of a study to clarify the features and advantages of internal anti-rotation configurations, the objective of the present investigation was to compare and evaluate the marginal fit and size of microgap at the implant-abutment interface for several external and internal anti-rotation configurations. To this end, three internal connection and two external hex connection implant-abutment assemblies were examined in this study. The implant-abutment interface (I-A interface) was evaluated using three geometrical factors: vertical and horizontal discrepancies and size of microgap. Marginal fit and microgap size were measured by a scanning laser microscope. The I-A interface was also observed using a scanning electron microscope. Mean vertical discrepancy ranged from 22.6 to 62.2 microm, while horizontal discrepancy ranged from -27.1 to 16.0 microm. The microgap values of all I-A interfaces assessed in this study ranged from 2.3 to 5.6 microm. In conclusion, SEM images of I-A interface suggested no relationship between the geometrical factors and the type of anti-rotation configuration.