Related Experiment Video
Updated: Jul 7, 2026

Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
Device dislocation, probably due to paroxysmal coughing early after percutaneous closure of secundum type atrial
Thomas Bartel1, Johannes O Bonatti, Silvana Müller
1Division of Cardiology, Department of Internal Medicine, Medical University Innsbruck, Innsbruck, Austria. thomas.bartel@i-med.ac.at
Abstract:
In a 56-year-old woman with a secundum-type atrial septal defect, a closure device was successfully implanted with primarily adequate positioning of the occluder. Two weeks after discharge, the patient developed paroxysmal coughing that probably led to a device dislocation. Months later, a considerable left-to-right shunt was detected, and the device was found to be ingrown in a defective position and had to be removed surgically. In conclusion, the case demonstrates that recurrent coughing may interfere with closure devices and may lead to dangerous dislocation.