Errors involved in the existing B-term expressions for the longitudinal diffusion in fully porous chromatographic
G Desmet1, K Broeckhoven, J De Smet
1Vrije Universiteit Brussel, Department of Chemical Engineering, Pleinlaan 2, 1050 Brussels, Belgium. gedesmet@vub.ac.be
Abstract:
Numerically solving the effective diffusion in a simplified representation of a chromatographic bed, it was found that the B-term expressions that have up to now been used in the literature, and which can all be reduced to either Deff=(gamma mDm+k'gamma sDs)/(1+k') or Deff=(gamma meDm+k''Dpart)/(1+k''), can no longer be considered to be unconditionally valid. This could be demonstrated by showing that the simulated diffusion data are in agreement with the mathematically sound effective medium theory (EMT), whereas the B-term expressions used up to now in literature are in conflict with the EMT, a theory that is widely accepted in all other fields of science. It is also shown that the use of the existing B-term expressions can lead to very large measurement errors (up to a 100% and more) for the determination of the stationary phase diffusion coefficient gamma sDs from peak parking experiments. The representation of the B-term diffusion should in the future hence be based on the Deff expressions that can be derived from the EMT. These are physically sound and are also more accurate than the classical B-term expressions used up to now.
More Related Videos
Related Concept Videos
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Column Efficiency: Plate Theory
A higher number of theoretical plates signifies better column efficiency and improved separation capabilities. Plate height affects bandwidth and separation quality; it is inversely...
Principles Of Column Chromatography
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Chromatography: Introduction
The phase in which the compounds linger or on which the compounds adsorb is called the stationary phase, whereas the mobile phase is the solvent that carries the solutes to be analyzed. In traditional column chromatography, the mixture flows through the stationary phase, and the compounds partition between the stationary and mobile phases...