Related Experiment Video
Updated: Jul 6, 2026

Compact Lens-less Digital Holographic Microscope for MEMS Inspection and Characterization
Published on: July 5, 2016
Simultaneous amplitude-contrast and quantitative phase-contrast microscopy by numerical reconstruction of Fresnel
E Cuche1, P Marquet, C Depeursinge
1Institute of Applied Optics, Swiss Federal Institute of Technology, CH-1015 Lausanne, Switzerland. etienne.cuche@epfl.ch
Abstract:
We present a digital method for holographic microscopy involving a CCD camera as a recording device. Off-axis holograms recorded with a magnified image of microscopic objects are numerically reconstructed in amplitude and phase by calculation of scalar diffraction in the Fresnel approximation. For phase-contrast imaging the reconstruction method involves the computation of a digital replica of the reference wave. A digital method for the correction of the phase aberrations is presented. We present a detailed description of the reconstruction procedure and show that the transverse resolution is equal to the diffraction limit of the imaging system.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Confocal Fluorescence Microscopy
Super-resolution Fluorescence Microscopy