Related Experiment Video
Updated: Jul 6, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Configuration of anchorage holes affects cemented fixation of the acetabular component in total hip replacement: an
Rajshree Mootanah1, Paul Jarrett, Paul Ingle
1Anglia Ruskin University, Chelmsford, Essex, UK. r.mootanah@anglia.ac.uk
Abstract:
Although surgical fixation techniques are major contributing factors to the survivorship of total hip replacements, they vary considerably among orthopaedic surgeons. We investigated the effect of the following configuration of anchorage holes on the stability of acetabular component fixation: 3 x 12 mm, 3 x 6 mm, 6 x 6 mm, and 12 x 6 mm. The reconstructed acetabulae were tested to torque failure, whilst being subjected to a compressive load of 2.1 KN. Higher torque to failure values were obtained for specimens with three 12 mm anchorage holes, compared with six or more 6 mm anchorage holes and were in line with our computer simulation results. We propose that the longevity of cemented total hip replacements could be improved by drilling a few large anchorage holes.