In vitro evaluation of the CeVOX continuous central venous oxygenation monitoring system
W Baulig1, A Dullenkopf, P Hasenclever
1Division of Cardiac Anaesthesia, University Hospital Zurich, Rämstrasse 100, 8091 Zurich, Switzerland. werner.baulig@usz.ch
Abstract:
We investigated the in vitro performance of the CeVOX system for continuous monitoring of central venous oxygen saturation by spectrophotometry (Pulsion Medical Systems, Munich, Germany). Oxygen inflow into the system was varied, and oxygen saturation values measured by CeVOX were documented. Blood samples were simultaneously taken to assess oxygen saturation by co-oximetry, and values were compared by Bland-Altman analysis. Sixty-six data pairs were obtained at CeVOX and co-oximetry values of 16-99% and 5.5-100%, respectively. Overall, CeVOX values only slightly overestimated co-oximetry values (mean bias +2.4%), but limits of agreement (2 SD of bias) were wide (-11.8 to +16.6%). Saturation measured by CeVOX underestimated that measured by co-oximetry at higher oxygen concentrations and overestimated it at lower oxygen concentrations. There was a nearly linear correlation of the mean bias, suggesting a systematic error. We conclude that the current version of the CeVOX system does not reliably reflect oxygen saturation.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.