Related Experiment Video
Updated: Jul 6, 2026

Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
Biomechanical characteristics of nonbridging external fixators for distal radius fractures
Go Yamako1, Yoshinori Ishii, Yoshikazu Matsuda
1Venture Business Laboratory, Niigata University, Niigata City, Niigata, Japan. yamako@gs.niigata-u.ac.jp
Purpose:
Nonbridging external fixation is becoming popular for distal radius fractures, although its biomechanical characteristics have not been documented. This study evaluated the biomechanical characteristics of nonbridging external fixators for distal radius fractures.
Methods:
We tested 3 currently available nonbridging fixators (F-Wrist fixator, Hoffman II Compact, and Pennig Dynamic Wrist Fixator) and determined their relative stiffness under axial compression, torsion, and bending moments (dorsal, volar, radial, and ulnar aspects) using a uniform unstable distal radius fracture model. The contact pressure and its total load on the fracture plane were also measured to evaluate the mechanical stimuli at the stable fracture site using a pressure-sensitive conductive rubber sensor.
Results:
Differences were observed in the stiffness: the Pennig fixator was the stiffest, whereas the F-Wrist fixator was the least rigid. The total load transmitted from the wrist joint to the fracture plane depended on the fixator stiffness in axial compression.
Conclusions:
By determining the biomechanical characteristics of nonbridging external fixators, these data may help the clinician when deciding on a particular device for nonbridging external fixation.
More Related Videos
11:21Development of a Novel Internal Fixation Model for Rat Radial Fractures: Fracture Healing Assessment and Dorsal Root Ganglion Isolation
Published on: March 13, 2026
08:20A Reliable and Reproducible Critical-Sized Segmental Femoral Defect Model in Rats Stabilized with a Custom External Fixator
Published on: March 24, 2019
Related Concept Videos
Bones of the Upper Limb: Radius
The radius has a nail-shaped head, and a short...
Bones of the Upper Limb: Ulna
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...