Related Experiment Video
Updated: Jul 6, 2026

Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
Real-time MR-guided wire localization of breast lesions by using an open 1.0-T imager: initial experience
Axel Gossmann1, Christopher Bangard, Mathias Warm
1Department of Radiology, University of Cologne, Joseph-Stelzmann-Strasse 9, D-50924 Cologne, Germany. GossmannA@kliniken-koeln.de
Abstract:
The purpose of this study was to prospectively evaluate technique and time factors for real-time magnetic resonance (MR) imaging-guided wire localization of suspicious breast lesions by using an open 1.0-T MR imager. It was conducted with institutional review board approval; informed consent was given by patients. Needle placement was monitored in 30 women (mean age, 50.5 years; range, 28-70 years) by using a dynamic balanced gradient-echo (single-shot turbo field-echo [TFE]) sequence with a temporal resolution of 0.5 second. In all patients, the tip of the needle was clearly identified during placement. Consistent with balanced TFE (BTFE) imaging, diagnostic MR imaging after the interventional procedure confirmed that the hookwires were placed 0-6 mm (mean, 3.3 mm) from the target lesions. The total procedure time ranged from 16-36 minutes. Results show that real-time MR-guided wire localization permits correction of the needle position during placement and reduces the interventional procedure time.

