Published on: October 11, 2016
Beam shape effects on grating spectrometer resolution
1NASA-Ames Research Center, MS 245-6, Moffett Field, California 94035-1000, USA. erickson@cygnus.arc.nasa.gov
Abstract:
The collimated optical beam in a grating spectrometer may be circular or elliptical in cross section, so that different parts of the beam illuminate different numbers of grooves on the grating. Here we estimate the consequent loss in spectral resolution relative to that obtained with a beam that illuminates a fixed number of grooves. The effect reduces the intrinsic resolving power of the spectrometer by approximately 15%, exclusive of other contributions such as finite entrance-slit width.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...