Assessment of vertebral artery stents using 16-slice multi-detector row CT angiography in vivo evaluation: comparison
Won Jong Yoo1, Yeon Soo Lim, Kook Jin Ahn
1Department of Radiology, Holy Family Hospital, College of Medicine, The Catholic University of Korea, Wonmi-Gu, Bucheon, Gyeonggi, Republic of Korea.
Objectives:
To assess the lumen visibility of extracranial vertebral artery stents examined with 16-slice multi-detector row computed tomography (MDCT) angiography in vivo using a medium-smooth kernel (B30s) and a sharp kernel (B60s), and to compare these with digital subtraction angiography (DSA) after stent placement.
Methods:
Twenty stents from 20 patients (14 men, 6 women; mean age, 62.7+/-10.1 years) who underwent CT angiography (CTA) with 16-slice MDCT were retrospectively analyzed. In CT angiograms using a B30s and a B60s, the lumen diameters and CT attenuations of the stented vessels were measured three times by three observers, and artificial luminal narrowing (ALN) was calculated. To assess measurement reliability on CT angiograms, the intraclass correlation coefficient (ICC) was used. DSA served as the reference standard for the in-stent luminal measurements on CT angiography. The median interval between CT angiography and DSA was 1 day (range 1-10).
Results:
For interobserver reliability, intraclass correlation coefficients for the lumen diameters on CT angiograms with a B30s and a B60s were 0.90 and 0.96, respectively. The lumen diameters on CT angiograms using a B30s were consistently smaller than that on CT angiograms using a B60s (p<0.01). The mean ALN was 37+/-7% on CT angiograms using a B30s and 25+/-9% on CT angiograms using a B60s. The mean CT attenuation in in-stent lumen was 347+/-55HU on CT angiograms using a B30s and 295+/-46HU on CT angiograms using a B60s. The ALN and CT attenuation within the stented vessels between CT angiograms using a B30s and a B60s was significant (p<0.01).
Conclusions:
16-slice MDCT using a sharp kernel allows good visualization of the stented vessels and is useful in the assessment of vertebral artery stent patency after stent placement.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies VII: Vascular Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...