Comparison of standard-dose vs low-dose attenuation correction CT on image quality and positron emission tomographic
Vikram Krishnasetty1, Ali A Bonab, Alan J Fischman
1Department of Radiology, Massachusetts General Hospital, Boston, Massachusetts 02114, USA. vkrishnasetty@partners.org
Objective:
To determine if low-dose attenuation correction computed tomography (CTAC) 1) provides images with acceptable anatomic definition and noise compared with standard-dose CTAC and 2) provides acceptable positron emission tomographic attenuation correction.
Methods:
Positron emission tomography/computed tomography was performed on 78 patients. Forty-three patients underwent CTAC with tube current based on a standard weight-based scale. A second group of 35 patients underwent CTAC with tube current based on a low-dose weight-based scale. In a blinded review, two radiologists rated each examination for anatomic definition and image noise at 4 major anatomic levels using a 5-point scale. To evaluate for correct positron emission tomographic attenuation correction at the standard and reduced radiation doses on CTAC, water phantom studies using both imaging techniques were performed and compared.
Results:
Patients who underwent low-dose CTAC received a mean 60.6% reduction in radiation dose compared with those who underwent standard-dose CTAC (P < .0001). Low-dose CTAC demonstrated statistically significant poorer ratings for anatomic detail and noise at each of the 4 anatomic levels (P < .0001) compared with standard-dose CTAC. Scans were graded acceptable for diagnostic interpretation if scores for image noise and anatomic definition were greater than 3 at all anatomic levels. There was a significant difference between the number of acceptable diagnostic scans in the standard-dose group (88.4%) compared with the low-dose group (17.1%) (P < .0001). There was no statistical difference in attenuation correction values in low-dose and standard-dose attenuation correction maps of the water phantom.
Conclusion:
Low-dose CTAC significantly reduced the effective dose while providing optimal positron emission tomographic attenuation correction. However, because of decreased image quality, low-dose CTAC was not acceptable for diagnostic interpretation.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography