Reproducibility and validity of a handheld spirometer
Related Experiment Video
Updated: Jul 6, 2026

Employing the Forced Oscillation Technique for the Assessment of Respiratory Mechanics in Adults
Published on: February 9, 2022
R Graham Barr1, Kimberly J Stemple, Sonia Mesia-Vela
1Department of Medicine, Columbia University Medical Center, New York, NY 10032, USA. rgb9@columbia.edu
Background:
Handheld spirometers have several advantages over desktop spirometers, but worries persist regarding reproducibility and validity of data from handheld spirometers. We undertook an independent examination of the EasyOne handheld spirometer.
Methods:
The laboratory testing included reproducibility and validity testing with a waveform generator. We used standard American Thoracic Society waveforms for in-line testing, calibration adaptor testing, and testing during compression of the mouthpiece. The clinical testing involved repeated tests with 24 spirometry-naïve volunteers and comparison to spirometry results from laboratory (volume-sensing dry rolling seal) spirometer.
Results:
The EasyOne exceeded standard thresholds for acceptability with the American Thoracic Society waveforms. In-line testing yielded valid results from the EasyOne. Between the EasyOne and the reference spirometer readings the mean +/- SD difference was 0.03 +/- 0.23 L for forced vital capacity (FVC) and -0.06 +/- 0.09 L for forced expiratory volume in the first second (FEV(1)). The calibration adaptor showed no appreciable problems. Extreme compression of the mouthpiece reduced the measured values. In clinical testing the coefficients of variation and limits of agreement were, respectively, 3.3% and 0.24 L for FVC, 2.6% and 0.18 L for FEV(1), and 1.9% and 0.05 for the FEV(1)/FVC ratio. The EasyOne readings were lower than those from the reference spirometer; the differences were: -0.12 L for FVC, -0.17 L for FEV(1), and -0.02 for FEV(1)/FVC. The limits of agreement were within criteria for FVC but not for the FEV(1), possibly due to a training effect.
Conclusion:
The EasyOne spirometer yielded generally reproducible results that were generally valid, compared to the values from the laboratory spirometer. The use of the EasyOne in clinical, occupational, and research settings seems justified.
Related Concept Videos
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Assessment of blood pressure in brachial artery(two-step method)
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Reliability and Validity
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: