Updated: Jul 6, 2026

A TIRF Microscopy Technique for Real-time, Simultaneous Imaging of the TCR and its Associated Signaling Proteins
Published on: March 22, 2012
A new TDCR-LS counter using Channel photomultiplier tubes
C Ivan1, P Cassette, Maria Sahagia
1Horia Hulubei National Institute of R&D for Physics and Nuclear Engineering , Bucharest, RO-77125, Romania. ivan@nipne.ro
Abstract:
A new Triple to Double Coincidence Ratio (TDCR) liquid scintillation (LS) counter using recently available photodetectors, the Channel photomultiplier (CPM) tubes, was constructed and tested in the framework of a scientific cooperation between IFIN-HH and LNHB. The prototype LS counter uses 3CPM tubes arranged symmetrically in an optical chamber around a standard LS vial. The behavior of the prototype was first evaluated with a light emitting diode (LED) light pulser. The counter was then compared against a TDCR counter using conventional photomultiplier tubes, by measuring (55)Fe, (3)H, (63)Ni and (90)Sr/(90)Y LS sources prepared in commercial liquid scintillation cocktails. Although the observed detection efficiency was significantly lower than the one achieved with the traditional counter, we found a remarkable agreement on the activity determination using the two counters. Details on the prototype and the measurement results obtained are discussed in this paper.