Related Experiment Video
Updated: Jul 6, 2026

A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management
Published on: June 23, 2023
Absolute calibration of image plates for electrons at energy between 100 keV and 4 MeV
Hui Chen1, Norman L Back, Teresa Bartal
1Lawrence Livermore National Laboratory, Livermore, California 94550-9234, USA. chen33@llnl.gov
Abstract:
We measured the absolute response of image plate (Fuji BAS SR2040) for electrons at energies between 100 keV and 4 MeV using an electron spectrometer. The electron source was produced from a short pulse laser irradiated on solid density targets. This paper presents the calibration results of image plate photon stimulated luminescence per electron at this energy range. The Monte Carlo radiation transport code MCNPX results are also presented for three representative incident angles onto the image plates and corresponding electron energy depositions at these angles. These provide a complete set of tools that allows extraction of our absolute calibration to other spectrometer setting at this electron energy range.
Related Concept Videos
Scanning Electron Microscopy
Fundamental Principles
Accelerated...
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Overview
Electron Behavior
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Overview of Electron Microscopy