Related Experiment Video
Updated: Jul 6, 2026

An Experimental Protocol for Femtosecond NIR/UV - XUV Pump-Probe Experiments with Free-Electron Lasers
Published on: October 23, 2018
A new sample holder for laser-excited pump-probe magnetic measurements on a Focus photoelectron emission microscope
Jorge Miguel1, Matthias Bernien, Daniela Bayer
1Institut für Experimentalphysik, Freie Universität Berlin, Arnimallee 14, D-14195 Berlin, Germany. miguel@physik.fu-berlin.de
Abstract:
A custom-made Omicron-compatible sample holder for time-resolved photoelectron emission microscopy experiments is presented. It comprises a sample plate with four contacts that hosts a chip carrier where the semiconductor substrate is mounted. Covering the sample holder, a 6 mm diameter mask protects electrostatically the sample from the extractor lens voltage while keeping the imaging quality unperturbed. The improvements are a greater sample lifetime and the ability to withstand much higher currents in the stripline that provides the magnetic pulse to the magnetic microstructure.