Published on: May 27, 2016
Patient skin dose during neuroembolization by multiple-point measurement using a radiosensitive indicator
S Suzuki1, S Furui, Y Matsumaru
1Department of Radiology, Teikyo University School of Medicine, Tokyo, Japan. s-suzuki@med.teikyo-u.ac.jp
AJNR. American Journal of Neuroradiology
|April 5, 2008
Summary
Patient radiation exposure during neuroembolization is significant. Entrance skin doses (ESDs) in some cases exceeded thresholds for skin injuries, highlighting the need for radiation safety awareness.
Area of Science:
- Radiology
- Medical Physics
- Neurointerventional Surgery
Background:
- Neuroembolization procedures are increasingly common.
- Radiation exposure risks associated with neuroembolization are not widely recognized.
- Patient safety concerning radiation dose requires further investigation.
Purpose of the Study:
- To evaluate patient entrance skin dose (ESD) during neuroembolization procedures.
- To assess radiation exposure across multiple institutions performing neuroembolizations.
Main Methods:
- A study involving 103 neuroembolizations across 6 institutions.
- Patient ESDs were measured using caps with radiosensitive indicators.
- Clinical follow-up was conducted to monitor for scalp effects.
Main Results:
- Average fluoroscopic time was 67.1 minutes; average digital subtraction angiography frames were 883.
- Average maximum patient ESD was 1.9 Gy, with institutional averages ranging from 1.0 to 2.4 Gy.
- Radiation-induced epilation occurred in 6 patients.
Conclusions:
- Maximum ESDs during neuroembolization can surpass safe limits for skin injury.
- The findings underscore the importance of monitoring and managing radiation doses in neurointerventional procedures.