Related Experiment Video
Updated: Jul 6, 2026

Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Progressive collapse of PMMA-augmented vertebra: a report of three cases
1Department of Neurosurgery, Yonsei University College of Medicine, Seoul, Republic of Korea.
Abstract:
Vertebroplasty using polymethylmethacrylate (PMMA) for augmentation is accepted as a safe and effective treatment for vertebral compression fracture. However, various complications related to PMMA vertebroplasty have recently been reported. We experienced three cases with progressive collapse of PMMA-augmented vertebra. Collapse progressed after augmentation in cases where PMMA conglomerated without contiguous bone interdigitation. A high viscosity of the PMMA preparation and vertebral body cavitory lesion may play a role in progressive vertebral collapse. To avoid this complication, bone cement should be injected sufficiently and permeate to contiguous bone to create strong support and anchorage.