Related Experiment Video
Updated: Jul 6, 2026

07:26
A Simple Method for Automated Solid Phase Extraction of Water Samples for Immunological Analysis of Small Pollutants
Published on: January 1, 2016
Steroid hormone determination in water using an environmentally friendly membrane based extraction technique
Saioa Zorita1, Pär Hallgren, Lennart Mathiasson
1Division of Analytical Chemistry, University of Lund, P.O. Box 124, SE-22100 Lund, Sweden. saioa.zorita@analykem.lu.se
Journal of Chromatography. A
|April 9, 2008
Summary
This study presents a new method for detecting steroid hormones like 17beta-estradiol in water. The hollow-fibre microporous membrane liquid-liquid extraction (HF-MMLLE) coupled with GC-MS offers sensitive and efficient analysis.
Area of Science:
- Environmental Chemistry
- Analytical Chemistry
- Water Quality Analysis
Background:
- Steroid hormones are emerging contaminants in water bodies.
- Accurate determination of steroid hormones is crucial for environmental monitoring.
- Existing methods may lack sensitivity or require complex sample preparation.
Purpose of the Study:
- To develop and optimize a sensitive method for determining specific steroid hormones in tap and sewage water.
- To evaluate the efficiency of hollow-fibre microporous membrane liquid-liquid extraction (HF-MMLLE) for steroid hormone analysis.
- To establish method performance characteristics, including limits of detection and repeatability.
Main Methods:
- Utilized a hollow-fibre microporous membrane liquid-liquid extraction (HF-MMLLE) technique.
- Coupled HF-MMLLE with gas chromatography-mass spectrometry (GC-MS) for analysis.
- Investigated parameters such as organic solvent, pH, salt, and humic acid content.
Main Results:
- Achieved low nanogram per liter (ng/L) level determination of steroid hormones in water samples.
- Obtained high correlation coefficients (>0.995) for linear calibration plots.
- Demonstrated method limits of detection as low as 1.6 ng/L for 17beta-estradiol in sewage water.
- Reported enrichment factors exceeding 1400 after derivatization.
- Showcased good repeatability, with values better than 10% and 6% at different concentrations.
Conclusions:
- The developed HF-MMLLE-GC-MS method is highly effective for the sensitive determination of steroid hormones in environmental water samples.
- The method offers a significant improvement in sensitivity and efficiency for water quality monitoring.
- This technique provides a reliable tool for assessing steroid hormone contamination in tap and sewage water.