Experience with the EEC phantom and test procedures in two block detector scanners.

T J Spinks1

  • 1MRC Cyclotron Unit, Hammersmith Hospital, London, UK.

Medical Progress Through Technology
|January 1, 1991
PubMed
Summary

Performance of two positron emission tomography (PET) scanners was compared. The brain scanner showed higher sensitivity and efficiency, especially with septa retracted, demonstrating improved PET imaging capabilities.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Diagnosis and Management of Mild Traumatic Brain Injury in Children: A Systematic Review.

JAMA pediatrics·2018
Same author

Quantitative PET and SPECT performance characteristics of the Albira Trimodal pre-clinical tomograph.

Physics in medicine and biology·2014
Same author

Influence of smoking, hypertension, and sex on the phenotypic expression of familial intracranial aneurysms in siblings.

Neurosurgery·2001
Same author

Effects of 7 years of growth hormone replacement therapy in hypopituitary adults.

The Journal of clinical endocrinology and metabolism·2000
Same author

Physical characteristics of the ECAT EXACT3D positron tomograph.

Physics in medicine and biology·2000
Same author

Generation of myocardial factor images directly from the dynamic oxygen-15-water scan without use of an oxygen-15-carbon monoxide blood-pool scan.

Journal of nuclear medicine : official publication, Society of Nuclear Medicine·1998

Area of Science:

  • Medical Imaging
  • Nuclear Medicine
  • Physics

Background:

  • Positron emission tomography (PET) scanners are crucial for medical diagnostics.
  • Evaluating the performance of commercial PET scanners is essential for optimizing clinical applications.

Purpose of the Study:

  • To compare the physical performance of two modern commercial PET scanners.
  • To assess transaxial resolution, sensitivity, scatter fraction, and count rate performance.

Main Methods:

  • Utilized a phantom and standardized test procedures from the EEC Concerted Action.
  • Compared a whole-body scanner (8 rings, 102 cm diameter) and a brain scanner (16 rings, 76 cm diameter).
  • Evaluated performance with and without septa retraction for the brain scanner.

Related Experiment Videos

Main Results:

  • Transaxial resolution and recovery coefficients were similar due to equal detector widths.
  • Brain scanner exhibited 7% higher normalized volume sensitivity with septa.
  • Septa retraction increased brain scanner efficiency by approximately 5x.
  • Scatter fractions were lower in reconstructed images compared to sinograms.
  • Accurate dead time correction was observed up to 480 MBq.

Conclusions:

  • Both PET scanners demonstrated comparable transaxial resolution.
  • The brain scanner offers significantly higher sensitivity and efficiency, particularly with septa retracted.
  • Reconstruction algorithms effectively manage scatter, and dead time correction is reliable at high activities.