Related Experiment Video
Updated: Jul 6, 2026

Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans
Published on: August 28, 2018
Diagnostic performance of 16- and 64-section spiral CT for coronary artery bypass graft assessment: meta-analysis
Michèle Hamon1, Olivier Lepage, Patrizia Malagutti
1Departments of Radiology, University Hospital of Caen, Avenue Côte de Nacre 14033 Caen, Normandy, France. hamon-mi@chu-caen.fr
Purpose:
To perform a meta-analysis to evaluate the accuracy of 16- and 64-section spiral computed tomography (CT) to help assess coronary artery bypass grafts (CABGs).
Materials And Methods:
The MEDLINE, Cochrane library, and BioMed Central databases were searched for relevant original articles published up to May 2007. Major criteria for article inclusion were that it (a) used multisection CT as a diagnostic test for the assessment of significant lesions (occlusion or >50% stenosis) of CABG, (b) used a 16- or 64-section scanner, and (c) used coronary angiography as the reference standard. After data extraction, the analysis was performed according to a random-effects model. Between-study statistical heterogeneity was also assessed by using the Cochran Q chi(2) test.
Results:
Of 158 screened articles, 15 fulfilled all inclusion criteria. Graft assessability (including distal anastomosis) ranged from 78%-100% among all included studies (mean, 92.4%; 90% with 16- and 96% with 64-section CT; P < .001). Statistical heterogeneity was observed for specificity and positive likelihood ratio (LR), justifying the use of the random-effects model. The analysis, pooled from 15 studies (723 patients, 2023 CABGs), provided the following results for the assessment of graft obstruction (occlusion and >50% stenosis): sensitivity, 97.6% (95% confidence interval [CI]: 96%, 98.6%); specificity, 96.7% (95% CI: 95.6%, 97.5%); positive predictive value, 92.7% (95% CI: 90.5%, 94.6%); negative predictive value, 98.9% (95% CI: 98.2%, 99.4%); positive LR, 23.42 (95% CI: 13.69, 40.07); negative LR, 0.045 (95% CI: 0.028, 0.071); and diagnostic odds ratio, 780.32 (95% CI: 379.12, 1606.1).
Conclusion:
Multisection CT provided high accuracy for the evaluation of CABG obstruction in assessable conduits, and might be used as a noninvasive tool for the evaluation of suspected graft dysfunction in patients who are at high risk for complications from coronary angiography.
More Related Videos
06:57Semi-Automatic Graphical Tool for Measuring Coronary Artery Spatially Weighted Calcium Score from Gated Cardiac Computed Tomography Images
Published on: September 22, 2023
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Acute Coronary Syndrome III: Diagnostic Studies