Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

NMR Spectrometers: Resolution and Error Correction01:14

NMR Spectrometers: Resolution and Error Correction

When magnetic nuclei in a sample achieve resonance and undergo relaxation, the signal detected in NMR is an approximately exponential free induction decay. Fourier transform of an exponential decay yields a Lorentzian peak in the frequency domain. Lorentzian peaks in an NMR spectrum are defined by their amplitude, full width at half maximum, and position, where the peak width is governed by the spin-spin relaxation time alone. In real experiments, however, the applied magnetic field is rendered...
Double Resonance Techniques: Overview01:12

Double Resonance Techniques: Overview

Double resonance techniques in Nuclear Magnetic Resonance (NMR) spectroscopy involve the simultaneous application of two different frequencies or radiofrequency pulses to manipulate and observe two distinct nuclear spins. One important application of double resonance is spin decoupling, which selectively suppresses coupling with one type of nucleus while observing the NMR signal from another nucleus, simplifying the spectrum and enhancing resolution.
Spin decoupling is usually achieved by...
Mass Analyzers: Common Types01:19

Mass Analyzers: Common Types

The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

IPEM Topical Report: a national survey of MRI use for external beam radiotherapy treatment planning in the UK: 2025 IPEM survey results.

Physics in medicine and biology·2026
Same author

Descriptive analysis of MRI functional changes occurring during reduced dose radiotherapy for myxoid liposarcomas.

The British journal of radiology·2021
Same author

IPEM Topical Report: an international IPEM survey of MRI use for external beam radiotherapy treatment planning.

Physics in medicine and biology·2021
Same author

An evaluation of systematic errors on marker-based registration of computed tomography and magnetic resonance images of the liver.

Physics and imaging in radiation oncology·2021
Same author

IPEM topical report: guidance on the use of MRI for external beam radiotherapy treatment planning<sup></sup>.

Physics in medicine and biology·2021
Same author

IPEM Topical Report: A 2018 IPEM survey of MRI use for external beam radiotherapy treatment planning in the UK.

Physics in medicine and biology·2019

Related Experiment Video

Updated: Jul 6, 2026

High-precision Electromagnetic Flowmeter with Empty Pipe Detection via Complex Programmable Logic Device-based Waveform Recognition
05:11

High-precision Electromagnetic Flowmeter with Empty Pipe Detection via Complex Programmable Logic Device-based Waveform Recognition

Published on: June 27, 2025

Two-point Dixon fat-water separation: improving reliability and accuracy in phase correction algorithms.

Maria A Schmidt1, Karlene M Fraser

  • 1Department of Medical Physics, St George's Hospital, London, UK. maria.schmidt@icr.ac.uk

Journal of Magnetic Resonance Imaging : JMRI
|April 15, 2008
PubMed
Summary

The advanced VSM algorithm offers improved fat-water separation, reducing errors in parotid MRI. This phase-correction technique enhances accuracy and speed for broader clinical use.

More Related Videos

Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol
07:59

Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol

Published on: September 7, 2018

A 3D Quantification Technique for Liver Fat Fraction Distribution Analysis Using Dixon Magnetic Resonance Imaging
05:37

A 3D Quantification Technique for Liver Fat Fraction Distribution Analysis Using Dixon Magnetic Resonance Imaging

Published on: October 20, 2023

Related Experiment Videos

Last Updated: Jul 6, 2026

High-precision Electromagnetic Flowmeter with Empty Pipe Detection via Complex Programmable Logic Device-based Waveform Recognition
05:11

High-precision Electromagnetic Flowmeter with Empty Pipe Detection via Complex Programmable Logic Device-based Waveform Recognition

Published on: June 27, 2025

Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol
07:59

Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol

Published on: September 7, 2018

A 3D Quantification Technique for Liver Fat Fraction Distribution Analysis Using Dixon Magnetic Resonance Imaging
05:37

A 3D Quantification Technique for Liver Fat Fraction Distribution Analysis Using Dixon Magnetic Resonance Imaging

Published on: October 20, 2023

Area of Science:

  • Medical Imaging
  • Biomedical Engineering
  • Magnetic Resonance Imaging

Background:

  • Accurate fat-water separation is crucial for MRI analysis, particularly in parotid gland imaging.
  • Existing phase-correction algorithms face challenges due to the complex spectral nature of biological tissues.

Purpose of the Study:

  • To introduce an advanced phase-correction region-growing algorithm (VSM) for improved two-point fat-water separation.
  • To evaluate the performance of the VSM algorithm against the original phase-correction (OPC) algorithm in parotid assessment.

Main Methods:

  • The study compared the original phase-correction (OPC) algorithm with a novel voxel size manipulation (VSM) algorithm.
  • VSM involves a multi-resolution region-growing approach and targeted low-pass filtering at fat-water interfaces.

Main Results:

  • The VSM algorithm demonstrated reduced noise sensitivity and increased speed compared to OPC.
  • In head images, VSM yielded significantly fewer errors (6) than OPC (43), with OPC specifically causing errors around the parotids.

Conclusions:

  • The VSM algorithm provides more accurate and robust fat-water separation, enhancing MRI analysis.
  • This improved performance broadens the applicability of phase-correction techniques in clinical settings.