Related Experiment Video
Updated: Jul 5, 2026

Closure of a Patent Foramen Ovale (PFO): An Intervention Sequence
Published on: December 23, 2022
Feasibility of patent foramen ovale closure with no-device left behind: first-in-man percutaneous suture closure
Carlos E Ruiz1, Nicholas Kipshidze, Paul T L Chiam
1Department of Cardiac and Vascular Interventional Services, Lenox Hill Heart and Vascular Institute of New York, New York, USA. cruiz@lenoxhill.net
Objectives:
This study evaluated the feasibility of percutaneous patent foramen ovale (PFO) closure using a transcatheter suture (Superstitch), leaving no device behind.
Background:
PFO has been implicated in cryptogenic strokes and migraine with auras. Percutaneous PFO closure, being less invasive than surgical closure, is increasingly performed. There are, however, early and long-term risks including device embolization, fracture, thrombosis, or infection, erosions into the free atrial wall and aorta, arrhythmias, and death. Furthermore, device implantation may complicate future percutaneous access to the left atrium. Partially reabsorbable devices and tissue welding to close PFO have recently been introduced. The first-in-man transcatheter suture closure of a PFO in an 18-year-old female with chronic migraine with aura and a well documented stroke is described.
Methods:
The right femoral vein was cannulated under mild sedation and local anesthesia. Using intracardiac echocardiography imaging, bubble study demonstrated a right-to-left shunt through the PFO at rest. A Superstitch device was advanced across the PFO and sutures were delivered through the septum primum and secundum. The sutures were exteriorized and a knot was advanced to the right atrial septum and cut.
Results:
Bubble study confirmed successful PFO suture closure. Transesophageal and transthoracic echocardiograms with bubble studies at 1 and 2 months, respectively showed complete closure with no right-to-left shunt even during Valsalva maneuver. At 6 months, the patient remained free of symptoms or migraine episodes.
Conclusion:
Percutaneous transcatheter suture closure of a PFO can be successfully achieved with no residual shunt and leaving no device behind. Technological refinements are required for wider scale use.
More Related Videos
08:12Fully Endoscopic Mitral Valve Repair with Percutaneous Cannulation of Groin Vessels
Published on: May 26, 2023
20:33A Case Series of Successful Abdominal Closure Utilizing a Novel Technique Combining a Mechanical Closure System with a Biologic Xenograft that Accelerates Wound Healing
Published on: July 4, 2019