Acoustic emissions during 3.1 MHz ultrasound bulk ablation in vitro
T Douglas Mast1, Vasant A Salgaonkar, Chandrapriya Karunakaran
1Department of Biomedical Engineering, University of Cincinnati, Cincinnati, OH 45267-0586, USA. doug.mast@uc.edu
Abstract:
Acoustic emissions associated with cavitation and other bubble activity have previously been observed during ultrasound (US) ablation experiments. Because detectable bubble activity may be related to temperature, tissue state and sonication characteristics, these acoustic emissions are potentially useful for monitoring and control of US ablation. To investigate these relationships, US ablation experiments were performed with simultaneous measurements of acoustic emissions, tissue echogenicity and tissue temperature on fresh bovine liver. Ex vivo tissue was exposed to 0.9-3.3-s bursts of unfocused, continuous-wave, 3.10-MHz US from a miniaturized 32-element array, which performed B-scan imaging with the same piezoelectric elements during brief quiescent periods. Exposures used pressure amplitudes of 0.8-1.4 MPa for exposure times of 6-20 min, sufficient to achieve significant thermal coagulation in all cases. Acoustic emissions received by a 1-MHz, unfocused passive cavitation detector, beamformed A-line signals acquired by the array, and tissue temperature detected by a needle thermocouple were sampled 0.3-1.1 times per second. Tissue echogenicity was quantified by the backscattered echo energy from a fixed region-of-interest within the treated zone. Acoustic emission levels were quantified from the spectra of signals measured by the passive cavitation detector, including subharmonic signal components at 1.55 MHz, broadband signal components within the band 0.3-1.1 MHz and low-frequency components within the band 10-30 kHz. Tissue ablation rates, defined as the thermally ablated volumes per unit time, were assessed by quantitative analysis of digitally imaged, macroscopic tissue sections. Correlation analysis was performed among the averaged and time-dependent acoustic emissions in each band considered, B-mode tissue echogenicity, tissue temperature and ablation rate. Ablation rate correlated significantly with broadband and low-frequency emissions, but was uncorrelated with subharmonic emissions. Subharmonic emissions were found to depend strongly on temperature in a nonlinear manner, with significant emissions occurring within different temperature ranges for each sonication amplitude. These results suggest potential roles for passive detection of acoustic emissions in guidance and control of bulk US ablation treatments.
More Related Videos
08:19Controllable Nucleation of Cavitation from Plasmonic Gold Nanoparticles for Enhancing High Intensity Focused Ultrasound Applications
Published on: October 5, 2018
07:38Real-time Monitoring of High Intensity Focused Ultrasound (HIFU) Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound (HMIFU)
Published on: November 3, 2015
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...