Related Experiment Video
Updated: Jul 5, 2026

Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Carbon dioxide rebreathing during non-invasive ventilation delivered by helmet: a bench study
Francesco Mojoli1, Giorgio A Iotti, Maddalena Gerletti
1Servizio di Anestesia e Rianimazione I, Fondazione IRCCS Policlinico S. Matteo, Piazzale Golgi 2, Pavia, Italy. f.mojoli@smatteo.pv.it
Objective:
To define how to monitor and limit CO(2) rebreathing during helmet ventilation.
Design:
Physical model study.
Setting:
Laboratory in a university teaching hospital.
Interventions:
We applied pressure-control ventilation to a helmet mounted on a physical model. In series 1 we increased CO(2) production (V'CO(2)) from 100 to 550 ml/min and compared mean inhaled CO(2) (iCO(2),mean) with end-inspiratory CO(2) at airway opening (eiCO(2)), end-tidal CO(2) at Y-piece (yCO(2)) and mean CO(2) inside the helmet (hCO(2)). In series 2 we observed, at constant V'CO(2), effects on CO(2) rebreathing of inspiratory pressure, respiratory mechanics, the inflation of cushions inside the helmet and the addition of a flow-by.
Measurements And Results:
In series 1, iCO(2),mean linearly related to V'CO(2). The best estimate of CO(2) rebreathing was provided by hCO(2): differences between iCO(2),mean and hCO(2), yCO(2) and eiCO(2) were 0.0+/-0.1, 0.4+/-0.2 and -1.3+/-0.5%. In series 2, hCO(2) inversely related to the total ventilation (MVtotal) delivered to the helmet-patient unit. The increase in inspiratory pressure significantly increased MVtotal and lowered hCO(2). The low lung compliance halved the patient:helmet ventilation ratio but led to minor changes in MVtotal and hCO(2). Cushion inflation, although it decreased the helmet's internal volume by 33%, did not affect rebreathing. A 8-l/min flow-by effectively decreased hCO(2).
Conclusions:
During helmet ventilation, rebreathing can be assessed by measuring hCO(2) or yCO(2), but not eiCO(2). It is directly related to V'CO(2), inversely related to MVtotal and can be lowered by increasing inspiratory pressure or adding a flow-by.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Acute Respiratory Failure-III
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...