High-frequency amplification with ITC-HA and BTE-HA. A comparative investigation
C von Buchwald1, F Pedersen, A Parving
1Department of Audiology, Bispebjerg Hospital, Copenhagen, Denmark.
Scandinavian Audiology
|January 1, 1991
Summary
In-the-canal hearing aids (ITC-HA) provide greater high-frequency amplification than behind-the-ear hearing aids (BTE-HA). While not improving speech recognition, patients preferred ITC-HAs, leading to higher continuous usage.
Area of Science:
- Audiology
- Hearing Aid Technology
- Speech and Hearing Sciences
Background:
- High-frequency hearing loss is common and significantly impacts communication.
- Conventional behind-the-ear hearing aids (BTE-HA) with open molds may not provide sufficient high-frequency amplification.
- In-the-canal hearing aids (ITC-HA) offer a potentially more discreet and effective amplification solution.
Purpose of the Study:
- To quantify the high-frequency amplification of an ITC-HA.
- To compare the amplification of an ITC-HA with a conventional BTE-HA using an open mold.
- To assess patient preference and long-term usage of ITC-HA versus BTE-HA.
Main Methods:
- Insertion gain measurements were performed on 22 subjects with bilateral high-frequency hearing loss.
- Participants were fitted with both an ITC-HA and a BTE-HA with an open mold.
- Subjective preference and continuous usage rates were recorded after a one-year observation period.
Main Results:
- The ITC-HA provided significantly greater high-frequency amplification compared to the BTE-HA.
- Neither hearing aid achieved amplification levels suggested by the POGO strategy.
- Despite superior amplification, the ITC-HA did not yield significantly better speech recognition scores than the BTE-HA.
- All subjects preferred the ITC-HA at fitting, and 68% used it continuously after one year.
Conclusions:
- ITC-HAs are a viable option for patients with significant high-frequency hearing loss.
- Patients demonstrate a strong preference for ITC-HAs over conventional BTE-HAs.
- The preference for ITC-HAs correlates with higher long-term usage rates, suggesting improved patient satisfaction.