Related Experiment Video
Updated: Jul 5, 2026

Extraction and Visualization of Protein Aggregates after Treatment of Escherichia coli with a Proteotoxic Stressor
Published on: June 29, 2021
Common excipients impair detection of protein aggregates during sedimentation velocity analytical ultracentrifugation
John P Gabrielson1, Kelly K Arthur, Brent S Kendrick
1Analytical Sciences, Amgen, Inc., 4000 Nelson Rd., Longmont, Colorado 80503, USA.
Abstract:
The final formulations of modern pharmaceutical protein products typically contain sugars or sugar alcohols as stabilizers. Migration of these sugars under the influence of an applied gravitational field during sedimentation velocity analytical ultracentrifugation (SV-AUC) produces dynamic density and viscosity gradients. If the formation of such gradients is not taken into account during data analysis, the capability of the SV-AUC technique to detect protein oligomers/aggregates may be dramatically impacted. In the example described here, the limit of quantitation (LOQ) of a simulated monoclonal antibody (mAb) dimer increases from 0.8% to 2.4% upon addition of 5% sorbitol to the formulation. This study uses simulated and experimental SV-AUC data to demonstrate the detrimental effect of dynamic gradients; it further explores how sophisticated data analysis techniques, including SEDFIT's inhomogeneous solvent options, may be used to mitigate the detection problems caused by the sedimentation of excipients.
Related Concept Videos
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...
Centrifugation
SDS-PAGE
A variation of gel electrophoresis, termed polyacrylamide gel electrophoresis (PAGE), is commonly used for separating proteins according to their molecular size by passing them through a polyacrylamide gel. Because of the varying charges associated with amino acid side chains, PAGE can be used to separate intact proteins...