Related Experiment Video
Updated: Jul 5, 2026

An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
A comparison of two techniques for knee arthrodesis: the custom made intramedullary Mayday nail versus a monoaxial
David Yeoh1, Richard Goddard, Paul Macnamara
1Orthopaedic Department, Conquest Hospital, East Sussex Hospitals NHS Trust, UK. david_yeoh@yahoo.co.uk
Abstract:
The most common indication for knee arthrodesis is pain and instability in an unreconstructable knee following an infected knee arthroplasty. In this study, we compare the use of the Mayday arthrodesis nail (Orthodynamics, Christchurch, UK) versus external fixation, Orthofix (Berkshire UK) and Stryker Hoffman II (County Cork, Ireland). All patients in this study underwent arthrodesis between 1995 and 2006 at Conquest Hospital, Hastings. In group A, 11 patients underwent arthrodesis with a Mayday nail. In all cases, the indications were infected total knee replacements (TKR). Three of these patients previously had failed attempts at arthrodesis with external fixation devices. In group B, seven patients underwent arthrodesis using external fixation. In six patients, the indication was infected TKRs. Results were reviewed retrospectively, with union assessed both clinically and radiologically. The mean inpatient stay for the Mayday nail group was 23 days (range 8-45 days) compared with 76 days (range 34-122) for the external fixation group (p<0.01, CI 95). Ten patients in group A went on to confirmed primary arthrodesis. One patient underwent revision arthrodesis with a Mayday nail and subsequently united. In group B only two patients achieved union. The rate of union was significantly greater in the Mayday nail group than the external fixation group (91% vs 29%, p<0.01). Of those patients that achieved union, there was no difference in the time to fusion between groups. Our study supported the existing literature and found that the Mayday nail appeared more effective than monoaxial external fixators for arthrodesis in the management of infected total knee replacements.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
05:34A Mini-Invasive Internal Fixation Technique for Studying Immobilization-Induced Knee Flexion Contracture in Rats
Published on: May 20, 2019