Reduction of patient-ventilator asynchrony by reducing tidal volume during pressure-support ventilation
Arnaud W Thille1, Belen Cabello, Fabrice Galia
1Henri Mondor Teaching Hospital, Paris XII University, Créteil, France. arnaud.thille@hmn.aphp.fr
Objective:
To identify ventilatory setting adjustments that improve patient-ventilator synchrony during pressure-support ventilation in ventilator-dependent patients by reducing ineffective triggering events without decreasing tolerance.
Design And Setting:
Prospective physiological study in a 13-bed medical intensive care unit in a university hospital.
Patients And Participants:
Twelve intubated patients with more than 10% of ineffective breaths while receiving pressure-support ventilation.
Interventions:
Flow, airway-pressure, esophageal-pressure, and gastric-pressure signals were used to measure patient inspiratory effort. To decrease ineffective triggering the following ventilator setting adjustments were randomly adjusted: pressure support reduction, insufflation time reduction, and change in end-expiratory pressure.
Measurements And Results:
Reducing pressure support from 20.0 cm H(2)O (IQR 19.5-20) to 13.0 (12.0-14.0) reduced tidal volume [10.2ml /kg predicted body weight (7.2-11.5) to 5.9 (4.9-6.7)] and minimized ineffective triggering events [45% of respiratory efforts (36-52) to 0% (0-7)], completely abolishing ineffective triggering in two-thirds of patients. The ventilator respiratory rate increased due to unmasked wasted efforts, with no changes in patient respiratory rate [26.5 breaths/min (23.1-31.9) vs. 29.4 (24.6-34.5)], patient effort, or arterial PCO(2). Shortening the insufflation time reduced ineffective triggering events and patient effort, while applying positive end-expiratory pressure had no influence on asynchrony.
Conclusions:
Markedly reducing pressure support or inspiratory duration to reach a tidal volume of about 6 ml/kg predicted body weight eliminated ineffective triggering in two-thirds of patients with weaning difficulties and a high percentage of ineffective efforts without inducing excessive work of breathing or modifying patient respiratory rate.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...