Related Experiment Video
Updated: Jul 5, 2026

Trabecular Meshwork Response to Pressure Elevation in the Living Human Eye
Published on: June 20, 2015
[Methods of measuring intraocular pressure independently of central corneal thickness]
1Augenabteilung, Asklepios Klinik Nord Heidberg, Tangstedter Landstrasse 400, 22415 Hamburg, Deutschland. a.hager@asklepios.com
Introduction:
Goldmann applanation tonometry (GAT) has been the gold standard for measuring intraocular pressure (IOP) for about 50 years. However, it depends on central corneal thickness (CCT) and is, therefore, prone to being incorrect. Dynamic contour tonometry (DCT) has recently been introduced to measure IOP independently of CCT; however, DCT is costly and difficult. IOP measurement using the ocular response analyzer (ORA) offers noncontact tonometry with declaration of the corneal-compensated IOP (IOPcc), which takes corneal hysteresis (CH) into account and is supposed to be independent of CCT.
Patients And Methods:
Using the ORA instrument, IOPcc was determined in 192 glaucoma eyes and 59 nonglaucoma eyes. Subsequently, measurement by DCT and GAT was performed. IOP measurements were compared and analyzed with respect to CCT and CH.
Results:
Average values were as follows: IOPcc, 18.38+/-6.3 mmHg; GAT, 14.69+/-4.5 mmHg; DCT, 15.17+/-3.9 mmHg; CH, 9.96+/-2.5 mmHg; CCT, 552+/-57 mum. Neither CCT nor CH differed between the two groups. There was a positive correlation between GAT and CCT that did not exist for IOPcc and DCT values. However, IOPcc and DCT differed significantly in Bland-Altman analysis (p<0.01). Furthermore, these two IOP values differed significantly with respect to CH and the level of IOP.
Conclusion:
Because IOPcc is not a primarily measured variable but also takes CH into account, a direct comparison of DCT and IOPcc values is not acceptable, and a simple correction factor may not be valid.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Glaucoma: Overview
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: