[Computed tomography: influence of varying tube current on patient dose and correctness of effective dose
V Hietschold1, A Koch, M Laniado
1Institut und Poliklinik für Radiolog. Diagnostik, Universitätsklinikum Carl-Gustav-Carus der TU Dresden. Volker.Hietschold@uniklinikum-dresden.de
Purpose:
Determination of the influence of tube currents varying during a CT scan on organ doses and on the effective dose as a function of patient constitution. Evaluation of the accuracy of effective dose calculations based on summarizing parameters (effective mAs, dose length product [DLP]) compared to calculations based on slice-specific tube currents.
Materials And Methods:
Investigation of the CT datasets of 806 patients acquired from the skull base to the proximal thigh with respect to the body mass index (BMI). The effective dose was calculated by means of slice-specific as well as region-specific conversion factors.
Results:
Dose optimization by means of variable tube current resulted in a reduction of the gonad dose in patients with BMI < or = 20 ... 21 kg/m (2) and of the effective dose in patients with BMI < or = 26 kg/m (2). Effective dose values calculated with the DLP for 90 % of the patients are within an interval of +/- 20 % of the values calculated using slice-specific tube currents.
Conclusion:
If tube current optimization during the CT scan was applied, for the scan region under investigation, at a BMI already below the German mean value, an increased effective dose was observed. Calculations of the effective dose on the basis of summarizing values such as DLP or effective mAs are of sufficient accuracy.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
10:23A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management
Published on: June 23, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET