A 12-year experience of bipolar steroid-eluting epicardial pacing leads in children

Maren Tomaske1, Bart Gerritse, Leo Kretzers

  • 1Division of Pediatric Cardiology, University Children's Hospital, Zurich, Switzerland. maren.tomaske@kispi.uzh.ch

Insights

Steroid-eluting bipolar epicardial leads show excellent performance in children, maintaining low pacing thresholds and favorable lead survival rates up to 12 years. These leads offer a reliable alternative for pediatric cardiac pacing needs.

Area of Science:

  • Pediatric Cardiology
  • Biomedical Engineering
  • Electrophysiology

Background:

  • Transvenous pacing is often unsuitable for children with cardiovascular abnormalities or small vasculature.
  • Epicardial lead implantation becomes a necessary alternative in these pediatric cases.
  • Steroid-eluting bipolar epicardial leads are evaluated for their efficacy and safety.

Purpose of the Study:

  • To assess the long-term performance of steroid-eluting bipolar epicardial pacing leads in a pediatric population.
  • To analyze pacing thresholds, sensing capabilities, and lead survival rates.
  • To determine the suitability of these leads for various pacing sites in children.

Main Methods:

  • Prospective enrollment of 114 children with 239 atrial and ventricular bipolar epicardial leads.
  • Follow-up duration up to 12.2 years (median 3.2 years).
  • Semi-annual lead data collection and analysis of pacing thresholds and sensing parameters.

Main Results:

  • Median pacing thresholds remained below 1.2 V at 0.5 ms for both atrial and ventricular leads.
  • No significant difference in thresholds between left/right atrial or left/right ventricular leads.
  • Superior sensing was observed for left ventricular leads compared to right ventricular leads (11.2 vs 7.7 mV).
  • Lead survival at 5 years was 94% for atrial and 85% for ventricular leads.

Conclusions:

  • Bipolar steroid-eluting epicardial leads provide excellent, persistent low pacing thresholds and sensing in children.
  • Favorable lead survival rates (85-94% at 5 years) support their use in growing, active pediatric patients.
  • These leads are a viable alternative for permanent pacing, including complex applications like resynchronization or defibrillator therapy.
Abstract