A novel selective-excitation RF pulse for magnetic resonance imaging
Liu Ying1, Li Jianqi, Li Gengying
1Shanghai Key Laboratory of Functional Magnetic Resonance Imaging, Department of Physics, East China Normal University, Shanghai 200062, People's Republic of China. ling2431@163.com
Object:
A selective-excitation radiofrequency (RF) pulse that uses hard pulses composed of a sequence of composite pulses with positive and negative phases (P/N pulse) is proposed herein. Because the amplitude of the RF signal is unchanged during the excitation, RF amplification can be accomplished using a nonlinear RF power amplifier (i.e., class C or D type).
Materials And Methods:
In this article, Fourier series have been first used to analyze the equivalence between the proposed P/N pulse and the conventional soft pulse on selective excitation. Subsequently, computer simulations based on density-matrix theory are used to compare the excitation profiles of both the soft and the P/N pulses.
Results:
The excitation profiles of the P/N pulses have been measured experimentally through a home-built 0.3-T magnetic resonance imaging (MRI) system. In addition, several slices of images have been obtained as proofs by using the multislice two-dimensional spin echo sequence through replacement of the conventional soft pulse by the proposed P/N pulse.
Conclusion:
Because the perfect selectivity of the proposed P/N pulse, it can be used for imaging studies to improve the efficiency of amplification at the lowest cost.