The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Computer assisted orthopedic surgery; its influence on prosthesis size in total knee replacement
H M J van der Linden-van der Zwaag1, R Wolterbeek, R G H H Nelissen
1Department Orthopaedic Surgery, Leiden University Medical Center, PO Box 9600, 2300 RC Leiden, The Netherlands. Enrike@vander-linden.nl
The Knee
|May 3, 2008
Summary
Computer Assisted Orthopaedic Surgery (CAOS) may lead to femoral component oversizing in Total Knee Arthroplasty (TKA). This study found a risk of oversizing the femoral component when using CAOS with the NexGen system.
Area of Science:
- Orthopaedic Surgery
- Medical Device Technology
- Surgical Navigation Systems
Background:
- Computer Assisted Orthopaedic Surgery (CAOS) has improved component alignment in Total Knee Arthroplasty (TKA).
- The accuracy of CAOS in determining component size remains less understood.
- Accurate component sizing is crucial for successful TKA outcomes.
Purpose of the Study:
- To evaluate the accuracy of component sizing (femoral and tibial) using the Brainlab CAOS system in TKAs.
- To compare CAOS-guided sizing with pre-operative templating and post-operative assessments.
- To identify potential risks associated with CAOS in component selection.
Main Methods:
- A comparative study of 40 Total Knee Arthroplasties (TKAs), with 20 using CAOS and 20 using conventional methods.
- Evaluation of femoral and tibial component sizes against pre-operative templating.
- Post-operative scoring to assess the adequateness of component size.
- Statistical analysis using kappa values to determine agreement and t-tests for significant differences.
Main Results:
- Fair agreement was observed between templated and implanted component sizes for both femur (kappa=0.38) and tibia (kappa=0.35).
- The CAOS group showed significantly higher rates of femoral component oversizing (p=0.020).
- No significant difference in tibial component sizing was found between CAOS and conventional groups.
Conclusions:
- There is an elevated risk of femoral component oversizing when utilizing the Brainlab CAOS system with the NexGen TKA components.
- CAOS accuracy in component sizing requires further investigation to optimize TKA procedures.
- Surgeons should be cautious regarding femoral component selection when employing CAOS in TKA.