Related Experiment Video
Updated: Jul 5, 2026

Chemiluminescence-based Assays for Detection of Nitric Oxide and its Derivatives from Autoxidation and Nitrosated Compounds
Published on: February 16, 2022
Flow-injection determination of cysteine in pharmaceuticals based on luminol-persulphate chemiluminescence detection
Amir Waseem1, Mohammad Yaqoob, Abdul Nabi
1Department of Chemistry, University of Balochistan, Quetta, Pakistan. waseemq2000@hotmail.com
Abstract:
A flow injection (FI) method is reported for the determination of l-cysteine, based on its enhancement on chemiluminescence (CL) emission of luminol oxidized by sodium persulphate in alkaline solution. The calibration graph was linear over the range 1.0 x 10(-9)-5.0 x 10(-7) mol/L (r(2) = 0.9992), with relative standard deviations (RSDs) in the range 1.1-2.3% (n = 4). The limit of detection (3 sigma blank) was 5.0 x 10(-10) mol/L with a sample throughput of 120/h. The method was applied to pharmaceuticals and the results obtained were in reasonable agreement with the amount labelled. The proposed method was also applied to cysteine in synthetic amino acid mixtures. Calibration graphs of N-acetylcysteine and glutathione over the range 1.0-50 x 10(-8) and 0.5-7.5 x 10(-7) mol/L were also established (r(2) = 0.998 and 0.9986) with RSDs in the range 1.0-2.0% (n = 4), and the limits of detection (3 sigma blank) were 5.0 x 10(-9) and 1.0 x 10(-8) mol/L, respectively.