High amplitude T-wave alternans precedes spontaneous ventricular tachycardia or fibrillation in ICD electrograms
Charles D Swerdlow1, Xiaohong Zhou, Olga Voroshilovsky
1Cedars Sinai Medical Center, Los Angeles, CA Medtronic Inc., Minneapolis, MN, USA. swerdlow@ucla.edu
Background:
T-wave alternans (TWA) increases acutely prior to ventricular tachycardia (VT) or fibrillation (VF) in animal studies, suggesting that it may provide a warning for VT/VF in implantable cardioverter defibrillator (ICD) patients. Clinically, measurement of surface ECG TWA requires preprocessing the input signal to reduce noise and/or analyzing more sinus beats than are recorded in ICDs as pre-onset, stored intracardiac electrograms (EGMs) before VT/VF. Our objective was to measure TWA from the few sinus EGMs stored in ICDs before spontaneous VT/VF in humans.
Objective:
The purpose of this study was to evaluate the technical feasibility of measuring TWA from pre-onset ICD EGMs and to measure EGM TWA before spontaneous VT/VF in humans.
Methods:
We developed a method to measure EGM TWA as a simple average (AVE) of peak-to-peak alternans. Using simulation, we determined the effect of ICD signal processing on EGM TWA for durations comparable to those in pre-onset EGMs. We then applied this method to pre-onset ICD EGMs that preceded 101 episodes of sustained VT/VF in 10 patients. In 6 of these patients, EGM recordings in atrial pacing and sinus rhythm provided control data.
Results:
In simulation, the AVE method discriminated input TWA differences > or = 15 microV. In patients, EGM TWA was 78 +/- 62 microV prior to VT/VF vs. 13 +/- 10 microV in control recordings (p< .0001). Eighty percent of pre-onset measurements exceeded 30 microV, while 95% of control measurements were less than 30 microV.
Conclusions:
A simple averaging method can measure TWA preceding VT/VF in stored ICD EGMs. Pilot data indicate that high-amplitude EGM TWA usually precedes spontaneous VT/VF and is infrequent in control recordings. They provide a rationale for developing ICD technology to measure EGM TWA continuously, both to warn patients and to initiate pacing algorithms to prevent VT/VF.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
09:17High-Resolution Endocardial and Epicardial Optical Mapping in a Sheep Model of Stretch-Induced Atrial Fibrillation
Published on: July 29, 2011
Related Concept Videos
Dysrhythmias III: Characteristics of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...