Anatomy-based inverse planning simulated annealing optimization in high-dose-rate prostate brachytherapy: significant
Dayee Jacob1, Adam Raben, Abhirup Sarkar
1Department of Radiation Oncology, Helen F. Graham Cancer Center, Christiana Care Health System, Newark, DE, USA.
Purpose:
To perform an independent validation of an anatomy-based inverse planning simulated annealing (IPSA) algorithm in obtaining superior target coverage and reducing the dose to the organs at risk.
Method And Materials:
In a recent prostate high-dose-rate brachytherapy protocol study by the Radiation Therapy Oncology Group (0321), our institution treated 20 patients between June 1, 2005 and November 30, 2006. These patients had received a high-dose-rate boost dose of 19 Gy to the prostate, in addition to an external beam radiotherapy dose of 45 Gy with intensity-modulated radiotherapy. Three-dimensional dosimetry was obtained for the following optimization schemes in the Plato Brachytherapy Planning System, version 14.3.2, using the same dose constraints for all the patients treated during this period: anatomy-based IPSA optimization, geometric optimization, and dose point optimization. Dose-volume histograms were generated for the planning target volume and organs at risk for each optimization method, from which the volume receiving at least 75% of the dose (V(75%)) for the rectum and bladder, volume receiving at least 125% of the dose (V(125%)) for the urethra, and total volume receiving the reference dose (V(100%)) and volume receiving 150% of the dose (V(150%)) for the planning target volume were determined. The dose homogeneity index and conformal index for the planning target volume for each optimization technique were compared.
Results:
Despite suboptimal needle position in some implants, the IPSA algorithm was able to comply with the tight Radiation Therapy Oncology Group dose constraints for 90% of the patients in this study. In contrast, the compliance was only 30% for dose point optimization and only 5% for geometric optimization.
Conclusions:
Anatomy-based IPSA optimization proved to be the superior technique and also the fastest for reducing the dose to the organs at risk without compromising the target coverage.