Film retakes in digital and conventional radiography
Waseem Akhtar1, Mubashir Aslam, Arif Ali
1Department of Radiology, The Aga Khan University Hospital, Karachi.
Objective:
To determine the film-retake rates and causes in digital radiography comparison to conventional X-rays method.
Study Design:
Comparative study.
Place And Duration Of Study:
Radiology Department, Aga Khan University Hospital, Karachi, from January 2004 to December 2006.
Patients And Methods:
X-rays of different body parts, conducted during the year 2004, with conventional radiography (n=170300), and in 2006 with digital radiography (n=174550), were included in this study. Measurements were done for number of X-rays re-take due to different quality control reasons for both the conventional and digital radiography. Quality control reasons included underexposure, overexposure, positioning errors, patient movements, portable X-rays, grid cutoff, and others (i.e. equipment related) due to which X-ray quality was questionable. Results were expressed in percentages.
Results:
A total of 9423 X-rays (5.5%) were repeated in conventional radiography (n=170300) due to underexposure (38%), overexposure (28.5%), positioning errors (25%), portable procedures (4%), patient movement (2%), grid cut-off (0.5%), and others (2%). Underexposure was the most frequently responsible factor for the X-ray repetition as compared to other factors (p<0.001). In digital radiography (n=174550), 1464 X-rays (1%) needed to be repeated, which was significantly less in comparison to X-ray repetition in conventional method of radiography (5.5%) [p<0.001]. In digital radiography, the most frequent factor for X-ray re-take was positioning error (435, 30%).
Conclusion:
Digital radiography is associated with significantly lesser number of re-take X-rays as compared to conventional radiography, hence minimizes the exposure of the patients to unnecessary radiations due to re-take X-rays. Positioning error remains a problem even in digital radiography, emphasizing training need for technologists.
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET