Related Experiment Video
Updated: Jul 5, 2026

An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
[Biomechanical study of a ventral stand-alone cage for the lumbar spine with and without additional posterior
1Abteilung für spezielle Wirbelsäulenchirurgie, Leopoldina-Krankenhaus, Schweinfurt, Germany.
Aim:
Biomechanical study to examine the stability of a stand-alone cage (SAC; Topaz, Ulrich Medizintechnik, Ulm) incorporating integrated anterior fixation with and without additional posterior fixation.
Method:
Six human spinal specimens (L4/5) were loaded in a spine tester in the three main motion planes in the following states: (a) intact, (b) defect (nucleotomy), (c) SAC, (d) SAC+internal fixator (IF), and (e) SAC+translaminar facet screws (TFS). Facet joint translation (FJT) and range of motion (ROM) were measured and used to evaluate the stability of the tested states.
Results:
The SAC stabilized the segment in comparison to the intact (a) and defect-containing (b) segments. The most rigid fixation was found for SAC+IF compared with the other states (a, b, c, e). The ROM and FJT of the SAC with the additional IF (d) showed a significant higher stability in all three motion planes. All differences in ROM and FJT between the tested states were statistically significant (p<0.05) except for the FJT SAC and SAC+TFS.
Conclusion:
The ventral Topaz SAC is a stable implant for the lumbar spine. Additional dorsal stabilization is an option to increase the stability.