Related Experiment Video
Updated: Jul 5, 2026

Laparoscopic Extracorporeal Knot-Tying for Uterine Vessel Occlusion during Hysterectomy with Cervical Cerclage in Large Uteri
Published on: September 12, 2025
Gynefix tales: cervical perforation and repeated late expulsion with the Gynefix device
Anjana Oswal1, Ashwini Oswal, Sofronis Loizides
1Family Planning and Reproductive Health, Luton, PCT, Luton, UK. oswal52@aol.com
Abstract:
We describe two unusual cases; both patients wish to continue using the Gynefix despite experiencing difficulty with the device. In case A, the marker string of the Gynefix(R) perforated the full thickness of the cervix and was visualized on the left lateral aspect of the cervix. Three months later the string had moved again and could no longer be visualized, but an ultrasound scan confirmed fundal implantation of the device. Case B demonstrates repeated late expulsion and failure of implantation of Gynefix. Appropriateness of ultrasound with reference to measurement of the distance SS (between the peritoneal surface of the uterine fundus and the first copper sleeve of the Gynefix), failure of implantation and problems with the new Gynefix introducer are discussed.
Related Concept Videos
Esophageal Perforation-I: Introduction
The location of esophageal perforation can vary, occurring anywhere along the esophagus.
Esophageal Perforation-II: Clinical Manifestations and Management
Clinical Manifestations:
Intrauterine Drug Delivery Systems