Related Experiment Video
Updated: Jul 5, 2026

Synthesis and Purification of Iodoaziridines Involving Quantitative Selection of the Optimal Stationary Phase for Chromatography
Published on: May 16, 2014
Determination of iodoacetic acid using liquid chromatography/electrospray tandem mass spectrometry
Benjamin P-Y Lau1, Adam Becalski
1Health Products and Food Branch, Health Canada, Ottawa, Ontario, Canada, K1A 0K9. ben_lau@hc-sc.gc.ca
Abstract:
A rapid analytical method based on liquid chromatography/tandem mass spectrometry (LC/MS/MS) using electrospray ionization in negative ion detection mode was developed for the analysis of underivatized iodoacetic acid in water. The method was applied to model reaction mixtures in the study of the formation of iodoacetic acid after chlorinated tap water was boiled in the presence of potassium iodide or iodized table salt. Samples can be directly analyzed by the LC/MS/MS system without extraction or chemical derivatization. Limit of detection was determined to be 0.3 microg/L (or 0.3 ng/mL) and limit of quantitation was about 1 microg/L (1 ng/mL).