The sound intensity and characteristics of variable-pitch pulse oximeters
Hiroo Yamanaka1, Junichi Haruna, Takashi Mashimo
1Department of Anesthesiology, Graduate School of Medicine, Osaka University, 2-2 Yamadaoka, Suita, Osaka, 565-0871, Japan. yhiroo@anes.med.osaka-u.ac.jp
Objective:
Various studies worldwide have found that sound levels in hospitals significantly exceed the World Health Organization (WHO) guidelines, and that this noise is associated with audible signals from various medical devices. The pulse oximeter is now widely used in health care; however the health effects associated with the noise from this equipment remain largely unclarified. Here, we analyzed the sounds of variable-pitch pulse oximeters, and discussed the possible associated risk of sleep disturbance, annoyance, and hearing loss.
Methods:
The Nellcor N 595 and Masimo SET Radical pulse oximeters were measured for equivalent continuous A-weighted sound pressure levels (L(Aeq)), loudness levels, and loudness. Pulse beep pitches were also identified using Fast Fourier Transform (FFT) analysis and compared with musical pitches as controls.
Results:
Almost all alarm sounds and pulse beeps from the instruments tested exceeded 30 dBA, a level that may induce sleep disturbance and annoyance. Several alarm sounds emitted by the pulse oximeters exceeded 70 dBA, which is known to induce hearing loss. The loudness of the alarm sound of each pulse oximeter did not change in proportion to the sound volume level. The pitch of each pulse beep did not correspond to musical pitch levels.
Conclusions:
The results indicate that sounds from pulse oximeters pose a potential risk of not only sleep disturbance and annoyance but also hearing loss, and that these sounds are unnatural for human auditory perception.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Regulation of Pulse
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...