A robust method for estimating regional pulmonary parameters in presence of noise
Richard A Guyer1, Michael D Hellman, Kiarash Emami
1Department of Radiology, University of Pennsylvania School of Medicine, 422 Curie Blvd, B1 Stellar-Chance Labs, Philadelphia, PA 19104-6100, USA.
Rationale And Objectives:
Estimation of regional lung function parameters from hyperpolarized gas magnetic resonance images can be very sensitive to presence of noise. Clustering pixels and averaging over the resulting groups is an effective method for reducing the effects of noise in these images, commonly performed by grouping proximal pixels together, thus creating large groups called "bins." This method has several drawbacks, primarily that it can group dissimilar pixels together, and it degrades spatial resolution. This study presents an improved approach to simplifying data via principal component analysis (PCA) when noise level prohibits a pixel-by-pixel treatment of data, by clustering them based on similarity to one another rather than spatial proximity. The application to this technique is demonstrated in measurements of regional lung oxygen tension using hyperpolarized (3)He magnetic resonance imaging (MRI).
Materials And Methods:
A synthetic dataset was generated from an experimental set of oxygen tension measurements by treating the experimentally derived parameters as "true" values, and then solving backwards to generate "noiseless" images. Artificial noise was added to the synthetic data, and both traditional binning and PCA-based clustering were performed. For both methods, the root-mean-square (RMS) error between each pixel's "estimated" and "true" parameters was computed and the resulting effects were compared.
Results:
At high signal-to-noise ratios (SNRs), clustering did not enhance accuracy. Clustering did, however, improve parameter estimations for moderate SNR values (below 100). For SNR values between 100 and 20, the PCA-based K-means clustering analysis yielded greater accuracy than Cartesian binning. In extreme cases (SNR<5), Cartesian binning can be more accurate.
Conclusions:
The reliability of parameters estimation in imaging-based regional functional measurements can be improved in the presence of noise by utilizing principal component analysis-based clustering without sacrificing spatial resolution compared to Cartesian binning. Results suggest that this approach has a great potential for robust grouping of pixels in hyperpolarized (3)He MRI maps of lung oxygen tension.
More Related Videos
05:56Phase-Resolved Functional Lung MRI for Pulmonary Ventilation and Perfusion (V/Q) Assessment
Published on: August 9, 2024
06:22Machine Learning-Based Cough Tone Classification: Diagnostic Exploration of Chronic Obstructive Pulmonary Disease and Respiratory Tract Infections
Published on: September 19, 2025
Related Concept Videos
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: