Related Experiment Video
Updated: Jul 5, 2026

In Vivo Luminal Measurement of Distension-Evoked Urothelial ATP Release in Rodents
Published on: September 7, 2022
Simulation of stress urinary incontinence for in-vitro studies
Florian Marti1, Nadine Blunschi, Thomas Leippold
1Computer Vision Laboratory, ETH Zürich, Switzerland.
Abstract:
A simulation system that generates dynamic bladder pressures for the use of testing and examining artificial urinary sphincters is designed, implemented, and compared to in-vivo measurements of Valsalva and coughing profiles. Cylinder and piston, which are integrated into the universal testing machine, simulating the bladder are connected with explanted sow urethras. The AMS 800 artificial urinary sphincter closes the urethra with well-defined external pressures. In order to select appropriate profiles for the bladder pressure, 34 Valsalva and coughing profiles of 6 patients were evaluated with respect to amplitude, pressure raise, dwell time, and half width.