A High Performance Impedance-based Platform for Evaporation Rate Detection
Published on: October 17, 2016
A new technique associated with the evaporimetry method for evaluating occlusion
J P P Marques1, D S B Basso, A S Nunes
1Natura Inovação e Tecnologia de Produtos Ltda., Rodovia Anhanguera, SP, Brazil. julianamarques@natura.net
International Journal of Cosmetic Science
|May 21, 2008
Summary
This study improved skin barrier assessment by combining evaporimetry with a product removal technique. This enhanced method accurately measures the occlusive effects of products like petroleum jelly on transepidermal water loss (TEWL).
Area of Science:
- Dermatology
- Biophysics
- Cosmetic Science
Background:
- The stratum corneum acts as a crucial skin barrier, regulating water loss and protecting against external agents.
- Maintaining skin hydration and elasticity relies on adequate lipid and water content.
- Evaporimetry is used to assess the occlusion potential of products that reduce transepidermal water loss (TEWL).
Purpose of the Study:
- To enhance the sensitivity of TEWL measurement for evaluating occlusive products.
- To compare the traditional evaporimetric method with a novel approach incorporating product removal.
Main Methods:
- The study compared traditional evaporimetric TEWL measurement with a new method combining evaporimetry and cotton-based occlusive product removal.
- Solid petrolatum was used as the occlusive product, applied to skin sites.
- TEWL measurements were taken before and after product removal, compared against untreated control sites.
Main Results:
- The traditional method failed to show a significant difference in occlusion between solid petrolatum and control sites.
- The enhanced method, including product removal, demonstrated a statistically significant difference in TEWL between occluded and control sites.
- Product removal significantly increased the sensitivity of evaporimetry for detecting occlusive effects.
Conclusions:
- The novel technique of combining evaporimetry with occlusive product removal significantly improves the sensitivity of TEWL measurements.
- This enhanced method provides a more accurate assessment of the occlusion potential of skincare products.
- The findings offer a more reliable approach for evaluating the efficacy of occlusive agents in dermatology and cosmetic science.